User Manual 2460 Series
Programming The 2461 Module 4-28
Racal Instruments
2003
Execution Errors
Error Description [Description/Explanation/Examples]
Number
-200 Execution Error
[This error occurs when the VXI output buffer is full, or if the INTERRUPT QUEUE is
full, or if the calibration date has been forgotten when closing the calibration process.]
-210 Trigger Error
-211 Trigger Ignored
[A *TRG or trigger signal has been received and recognized by the instrument but, for
timing considerations, has been ignored. For example, the instrument is not ready to
respond.]
-221 Configuration Conflict
[A command has been interpreted but cannot be executed because of the current state
of the instrument.]
Examples:
[ While the device is in INITiate state, some commands are not allowed.
A TTLTrg line is already forced in input or output.
An other TTLTrg line is already enabled in Output.
A CALCulation SCPI command is not allowed while Calculation state is OFF.]
-222 Data Out of Bounds
[A command has been interpreted but cannot be executed because its value is out of
bounds.]
-224 Illegal Parameter Value
[Used when a certain value, when possible, is expected. For example, the “CAL:STAT
HOLD” command generates an Error because the interpreter expects an ON or OFF
parameter instead of HOLD.]