User Manual 2460 Series
Programming The 2461 Module 4-12
Racal Instruments
2003
The sum of the decimal weights of the bits that you wish to set is sent as the
parameter ( <numeric_value> ) for the appropriate :ENABle: command.
Initialization:
*RST, *CLS
Not affected
Power-up:
0
Note that after power-up, by default, none of the bits of the
STATus:OPERation:EVENt register are reported to bit 7 of the Status Byte.
Examples:
"STAT:OPER:ENAB 512" enables the INIT bit of the
STATus:OPERation:EVENt
to
be reported in bit B7 of
the Status Byte.
Interrogative form:
STATus:OPERation:ENABle?
==>
<numeric_value>
*******************************************************************
STATus:OPERation:CONDition?
==> <numeric_value>
Description
:
Returns the contents of the Operation CONDition register.
Refer to Figure 4-2.
Content of the Status Operation Condition Register:
Bit B8, Testing state (Test): Set when the instrument has completed the self-test
process and clear when it starts the testing process.
Bit B9, Initiate state (Init1):
Set when the first instrument is in the Initiate state
(Trigger mode) and clear when it returns in the idle state (normal mode).
Bit B10, Data available (Meas1):
Set when all of the required measurements
have been completed by instrument 1. Cleared when the measurement is in
progress.