User Manual 2460 Series
Racal Instruments
2003
Programming The 2461 Module 4-11
STATus:PRESet
Description:
Clears to zero all bits of the Status Operation Enable Register.
Clears to zero all bits of the Status Operation Condition Register.
This command
does not erase the EVENt registers
.
**********************************************
STATus:OPERation[:EVENt]?
==> <numeric_value>
Description:
Returns the contents of the Operation Event register. Erases the contents of the
Operation Event register after reading.
Refer to Figure 4-2.
**********************************************
STATus:OPERation:ENABle <numeric_value>
Description:
Sets the specified ENABle register. Each enable register is used as a mask for
events. When a bit in an event enable register is cleared (0), the corresponding bit in
the event register is masked and thus, cannot set the corresponding bit of the next
register set in the status structure. Conversely, when a bit in an event enable
register is set (1), the corresponding bit in the event register is unmasked. When the
unmasked bit in the event register sets, the summary bit of the next register set in
the status structure will set.
Refer to Figure 4-2.
Parameter:
<numeric_value>
= 0
Clear enable register
= 256 Set bit B8 of enable register
= 512 Set bit B9 of enable register
= 1024 Set bit B10 of enable register
= 2048 Set bit B11 of enable register
= 4096 Set bit B12 of enable register
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com