RHYTHM SERVICE MANUAL
PAGE 3.1
2009
SUNRISE MEDICAL
Section 3
Check that the batteries are fully charged and in good
condition; and check all cables and connections. Check
the connections to the left motor, look for a loose or
damaged connector.
Use the meter to check the resistance across the two
bottom contacts (thicker wires) on the 4-pin motor
connector as shown in (figure 3.1.1). If the meter reads
between 0 to 1.5 ohms, then replace the controller. If
none of the above corrects the problem, replace the left
motor.
Two Bars - Left Motor Disconnected
One Bar - Low Battery Voltage )
This code could indicate discharged batteries, failed batteries, or poor battery connections. Begin by recharging
the batteries and then refer to Section 1 to check batteries and connections.
A3.1.2
A3.1.1
Otherwise, check the brushes on the left motor (Figure
A3.1.2). Ensure that they are not excessively worn.
Replace as required.
VR2 Controller Diagnostic Codes
Содержание Groove
Страница 1: ......
Страница 2: ...2 2006 Sunrise Medical ...
Страница 14: ...14 2006 Sunrise Medical INTRODUCTION ...
Страница 119: ...2009 Quickie Rhythm Service Manual Supplement 2009 Sunrise Medical Inc xxxxxx Rev A PG DRIVES TECHNOLOGY R ...
Страница 132: ...RHYTHM SERVICE MANUAL PAGE 0 14 2009 SUNRISE MEDICAL ...
Страница 138: ...RHYTHM SERVICE MANUAL 2009 PAGE 2 4 SUNRISE MEDICAL Section 2 ...
Страница 144: ...RHYTHM SERVICE MANUAL 2009 PAGE 3 6 SUNRISE MEDICAL ...
Страница 148: ...RHYTHM SERVICE MANUAL 2009 PAGE 4 4 SUNRISE MEDICAL ...
Страница 162: ...RHYTHM SERVICE MANUAL 2009 PAGE 6 6 SUNRISE MEDICAL ...
Страница 168: ...RHYTHM SERVICE MANUAL 2009 PAGE 8 4 SUNRISE MEDICAL Tilt to Interface Mounting Section 8 ...
Страница 179: ...RHYTHM SERVICE MANUAL PAGE 9 1 2009 SUNRISE MEDICAL Section 9 OBP Programming Quick Guide ...
Страница 180: ...RHYTHM SERVICE MANUAL 2009 PAGE 9 2 SUNRISE MEDICAL Section 9 OBP Programming Quick Guide ...
Страница 181: ......
Страница 182: ...182 2006 Sunrise Medical ...