RHYTHM SERVICE MANUAL
2009
PAGE 3.2
SUNRISE MEDICAL
Section 3
Check that the batteries are fully charged and in good
condition; and check all cables and connections. Check
the connections to the left motor, look for a loose or
damaged
connector.
Measure the resistance from the bottom contact of
the red thick wire on the 4-pin left motor connector to
each of the top contacts of the connector (figure 3.2.1).
Measure the resistance from the bottom contact of the
black thick wire on the 4-pin left motor connector to
each of the top contacts of the connector see (below
right). If all of the readings are open, then replace the
controller. If any of the readings are short, then replace
the left motor.
Three Bars - Left Motor Wiring Trip)
A3.2.1
Test 1
Test 2
Test 3
Test 4
VR2 Controller Diagnostics Codes (cont.)
Содержание Groove
Страница 1: ......
Страница 2: ...2 2006 Sunrise Medical ...
Страница 14: ...14 2006 Sunrise Medical INTRODUCTION ...
Страница 119: ...2009 Quickie Rhythm Service Manual Supplement 2009 Sunrise Medical Inc xxxxxx Rev A PG DRIVES TECHNOLOGY R ...
Страница 132: ...RHYTHM SERVICE MANUAL PAGE 0 14 2009 SUNRISE MEDICAL ...
Страница 138: ...RHYTHM SERVICE MANUAL 2009 PAGE 2 4 SUNRISE MEDICAL Section 2 ...
Страница 144: ...RHYTHM SERVICE MANUAL 2009 PAGE 3 6 SUNRISE MEDICAL ...
Страница 148: ...RHYTHM SERVICE MANUAL 2009 PAGE 4 4 SUNRISE MEDICAL ...
Страница 162: ...RHYTHM SERVICE MANUAL 2009 PAGE 6 6 SUNRISE MEDICAL ...
Страница 168: ...RHYTHM SERVICE MANUAL 2009 PAGE 8 4 SUNRISE MEDICAL Tilt to Interface Mounting Section 8 ...
Страница 179: ...RHYTHM SERVICE MANUAL PAGE 9 1 2009 SUNRISE MEDICAL Section 9 OBP Programming Quick Guide ...
Страница 180: ...RHYTHM SERVICE MANUAL 2009 PAGE 9 2 SUNRISE MEDICAL Section 9 OBP Programming Quick Guide ...
Страница 181: ......
Страница 182: ...182 2006 Sunrise Medical ...