If using the Quickie iQ, select Monitor Parameter
10.
followed by selecting Brakes. (
fi
g 2.23.15)
Operate the wheelchair by moving the joystick slight forward. Observe the Brake Current and Brake Voltage.
11.
Refer to the following notes:
Move the joystick further forward, followed by moving it to the left, then right, then reverse. Refer to the fol-
12.
The Rhythm and Groove use 12 VDC park brakes. Typical Brake Voltage is 11 VDC.
•
The Brake Current should be approximately 0.5 amps. Typical values are between 0.4 amps and
•
0.6 amps.
The Brake Voltage and Brake Current should remain constant.
•
An increase in Brake Current or Brake Voltage indicates that the park brake is shorted to the motor
•
windings.
lowing notes:
Fig 2.23.15
Содержание Groove
Страница 1: ......
Страница 2: ...2 2006 Sunrise Medical ...
Страница 14: ...14 2006 Sunrise Medical INTRODUCTION ...
Страница 119: ...2009 Quickie Rhythm Service Manual Supplement 2009 Sunrise Medical Inc xxxxxx Rev A PG DRIVES TECHNOLOGY R ...
Страница 132: ...RHYTHM SERVICE MANUAL PAGE 0 14 2009 SUNRISE MEDICAL ...
Страница 138: ...RHYTHM SERVICE MANUAL 2009 PAGE 2 4 SUNRISE MEDICAL Section 2 ...
Страница 144: ...RHYTHM SERVICE MANUAL 2009 PAGE 3 6 SUNRISE MEDICAL ...
Страница 148: ...RHYTHM SERVICE MANUAL 2009 PAGE 4 4 SUNRISE MEDICAL ...
Страница 162: ...RHYTHM SERVICE MANUAL 2009 PAGE 6 6 SUNRISE MEDICAL ...
Страница 168: ...RHYTHM SERVICE MANUAL 2009 PAGE 8 4 SUNRISE MEDICAL Tilt to Interface Mounting Section 8 ...
Страница 179: ...RHYTHM SERVICE MANUAL PAGE 9 1 2009 SUNRISE MEDICAL Section 9 OBP Programming Quick Guide ...
Страница 180: ...RHYTHM SERVICE MANUAL 2009 PAGE 9 2 SUNRISE MEDICAL Section 9 OBP Programming Quick Guide ...
Страница 181: ......
Страница 182: ...182 2006 Sunrise Medical ...