RHYTHM SERVICE MANUAL
PAGE 4.2
2009
SUNRISE MEDICAL
Section 4
5.
Disconnect the power connector from the
control module, and use a Multimeter to check
for battery voltage at the connector (fig4.2.2). If
voltage is present, replace the control module.
Retest as necessary.
Note: The power connector is the larger 2-pin
connector between the left and right motor
cable.
6.
Manually tilt the seat back and remove the
shroud from the base. Refer to Battery Removal
section of this manual.
7.
Disconnect the batteries (fig 4.2.3).
R-net Troubleshooting Procedures
fig. 4.2.2
fig. 4.2.1
fig. 4.2.3
4. Disconnect the joystick harness (bus cable) from
the controller. Use a Multimeter to test voltage at
the positive (+ red) and negative (- black)
ports in the controller module (
fi
g. 4.2.1). If voltage
is present, replace the following components in
this order:
a. Harness
b. Joystick
If voltage is NOT present please proceed to step
5.
Содержание Groove
Страница 1: ......
Страница 2: ...2 2006 Sunrise Medical ...
Страница 14: ...14 2006 Sunrise Medical INTRODUCTION ...
Страница 119: ...2009 Quickie Rhythm Service Manual Supplement 2009 Sunrise Medical Inc xxxxxx Rev A PG DRIVES TECHNOLOGY R ...
Страница 132: ...RHYTHM SERVICE MANUAL PAGE 0 14 2009 SUNRISE MEDICAL ...
Страница 138: ...RHYTHM SERVICE MANUAL 2009 PAGE 2 4 SUNRISE MEDICAL Section 2 ...
Страница 144: ...RHYTHM SERVICE MANUAL 2009 PAGE 3 6 SUNRISE MEDICAL ...
Страница 148: ...RHYTHM SERVICE MANUAL 2009 PAGE 4 4 SUNRISE MEDICAL ...
Страница 162: ...RHYTHM SERVICE MANUAL 2009 PAGE 6 6 SUNRISE MEDICAL ...
Страница 168: ...RHYTHM SERVICE MANUAL 2009 PAGE 8 4 SUNRISE MEDICAL Tilt to Interface Mounting Section 8 ...
Страница 179: ...RHYTHM SERVICE MANUAL PAGE 9 1 2009 SUNRISE MEDICAL Section 9 OBP Programming Quick Guide ...
Страница 180: ...RHYTHM SERVICE MANUAL 2009 PAGE 9 2 SUNRISE MEDICAL Section 9 OBP Programming Quick Guide ...
Страница 181: ......
Страница 182: ...182 2006 Sunrise Medical ...