66
©2006 Sunrise Medical
QUICKIE ELECTRONICS SECTION
Actuator Test (Motor Controller controlled actua-
tors)
Complete the following steps to test the seating system
actuators.
Connect the Quickie iQ or HHP to the wheel-
1.
chair.
Turn on the wheelchair. Prior to completing any
2.
test, re-home the wheelchair.
If using the HHP, select Monitor, then Motor Con-
3.
troller. Under Motor Controller, select the actuator
to test (i.e., Tilt). (
fi
g 2.23.22)
If using the Quickie iQ, select Monitor Parameter
4.
followed by Motor Controller, then actuator to test
(i.e., Actuator 1). (
fi
g 2.23.23)
Note the position of the actuator (counts). Operate the actuator and observe the counts and current. Refer to
5.
Table 1 for a listing of currents typical for each actuator.
Tilt
+ + -
Counts
0
Current
0.000
Fig 2.23.22
Fig 2.23.23
Содержание Groove
Страница 1: ......
Страница 2: ...2 2006 Sunrise Medical ...
Страница 14: ...14 2006 Sunrise Medical INTRODUCTION ...
Страница 119: ...2009 Quickie Rhythm Service Manual Supplement 2009 Sunrise Medical Inc xxxxxx Rev A PG DRIVES TECHNOLOGY R ...
Страница 132: ...RHYTHM SERVICE MANUAL PAGE 0 14 2009 SUNRISE MEDICAL ...
Страница 138: ...RHYTHM SERVICE MANUAL 2009 PAGE 2 4 SUNRISE MEDICAL Section 2 ...
Страница 144: ...RHYTHM SERVICE MANUAL 2009 PAGE 3 6 SUNRISE MEDICAL ...
Страница 148: ...RHYTHM SERVICE MANUAL 2009 PAGE 4 4 SUNRISE MEDICAL ...
Страница 162: ...RHYTHM SERVICE MANUAL 2009 PAGE 6 6 SUNRISE MEDICAL ...
Страница 168: ...RHYTHM SERVICE MANUAL 2009 PAGE 8 4 SUNRISE MEDICAL Tilt to Interface Mounting Section 8 ...
Страница 179: ...RHYTHM SERVICE MANUAL PAGE 9 1 2009 SUNRISE MEDICAL Section 9 OBP Programming Quick Guide ...
Страница 180: ...RHYTHM SERVICE MANUAL 2009 PAGE 9 2 SUNRISE MEDICAL Section 9 OBP Programming Quick Guide ...
Страница 181: ......
Страница 182: ...182 2006 Sunrise Medical ...