QUICKIE ELECTRONICS SECTION
Delphi Diagnostics Trouble Shooting Chart
Symptom
Code Flash
Diagnosis
Solution
Battery Voltage has
Dropped Lower or Raised
Higher Than Operating
Voltage
1 Bar – every 2 seconds
Batteries Depleted
Loose or Disconnected
battery
Bad Cells
Overcharged Batteries
Recharge batteries
Test with Volt Meter
Load Test Batteries
Check Connections
Replace batteries with New Set. Check Battery
Condition for Overcharged You will notice
sides of Batteries Concaved (overcharged)
Disconnect Left. Motor Con-
nection
2 bars – every 2 seconds
Motor 2 Open Circuit
Check motor connector
Check Armature Resistance
Switch Motor Leads
Replace Brushes
Replace Motor
Disconnect Right. Motor
Connection
3 bars – every 2 seconds
Motor 1 Open Circuit
Check motor connector
Check Armature Resistance
Switch Motor Leads
Replace Brushes
Replace Motor
Disconnect of Left or Right
Brake Connection or com-
plete Motor Connection
4 bars – every 2 seconds
Park Brake 1 or Park Brake
2 open circuit
Reconnect brake
Check brake resistance
Switch motor leads replace motor
Active Input Not Neutral At
Power-On Error
5 Bars – every 2 seconds
Active Input Not Neutral At
Power-On Error
Check to see if client has their hand on remote
Reset Power check integrity of Joystick Gimbal
If bent Replace out
Replace Joystick Module.
Loss of Communication
6 Bars – every 2 seconds
Loss of Communication
with Bus Line
Devices on Bus
Check wiring connections
Check all bus connections
Check diagnostic readings Replace compo-
nents causing failure if constant problem.
Motor Controller Internal
Module Error
7 Bars – every 2 seconds
Motor Controller Internal
Module Error
Replace Control module
Check all connections
Motor Controller High Tem-
perature Warning
1 Green (far right green)
Motor Controller High Tem-
perature Warning is TRUE
Stop Chair
Check Motor Connections
Check Motor resistance (Refer to Section 4)
Check all connections
Replace Control Module
Invalid System Con
fi
gura-
tion
2 Green (right to left)
Invalid System Con
fi
gura-
tion Error is TRUE
Check to see if Module is compatible with
system
If using QC control and hooking up non proto-
col device error will occur
If compatible with system try another control
module unplug Device
Drive Lockout External
Source
3 Green (right to left)
Drive Lockout Source is
True
Connection to source of inhibit such as tilt
system Recline
Charger is Plugged in
Ripples up from 1st LED to
Top LED #7
Charger Plugged In – In-
hibit Circuit
Disconnect Charger
Check wiring for Common Pin 2 and Pin 3
short
Содержание Groove
Страница 1: ......
Страница 2: ...2 2006 Sunrise Medical ...
Страница 14: ...14 2006 Sunrise Medical INTRODUCTION ...
Страница 119: ...2009 Quickie Rhythm Service Manual Supplement 2009 Sunrise Medical Inc xxxxxx Rev A PG DRIVES TECHNOLOGY R ...
Страница 132: ...RHYTHM SERVICE MANUAL PAGE 0 14 2009 SUNRISE MEDICAL ...
Страница 138: ...RHYTHM SERVICE MANUAL 2009 PAGE 2 4 SUNRISE MEDICAL Section 2 ...
Страница 144: ...RHYTHM SERVICE MANUAL 2009 PAGE 3 6 SUNRISE MEDICAL ...
Страница 148: ...RHYTHM SERVICE MANUAL 2009 PAGE 4 4 SUNRISE MEDICAL ...
Страница 162: ...RHYTHM SERVICE MANUAL 2009 PAGE 6 6 SUNRISE MEDICAL ...
Страница 168: ...RHYTHM SERVICE MANUAL 2009 PAGE 8 4 SUNRISE MEDICAL Tilt to Interface Mounting Section 8 ...
Страница 179: ...RHYTHM SERVICE MANUAL PAGE 9 1 2009 SUNRISE MEDICAL Section 9 OBP Programming Quick Guide ...
Страница 180: ...RHYTHM SERVICE MANUAL 2009 PAGE 9 2 SUNRISE MEDICAL Section 9 OBP Programming Quick Guide ...
Страница 181: ......
Страница 182: ...182 2006 Sunrise Medical ...