RHYTHM SERVICE MANUAL
PAGE 5.3
2009
SUNRISE MEDICAL
R-net Fault Codes cont.
Section 5
Bad Cable
Inspect and replace the defective cable(s).
Bad Settings
Verify that the programming agrees with the installed equipment. If all settings
are correct, the control module may be defective. Contact Sunrise Technical
Service for assistance.
Module Error
Recycle the power. Verify that the cables are correctly mated. If a module has
been replaced recently, suspect that module may be defective. Contact Sunrise
Technical Service for assistance.
System Error
Recycle the power. Verify that
the cables are correctly mated.
If a module has been replaced
recently, suspect that module
may be defective.
Contact Sunrise Technical
Service for assistance.
SID Detached
The Omni has detected that a specialty control has become disconnected.
Recheck all cables. If the error is still present, replace the specialty control.
User Switch
Detached
Indicates that a user switch has become disconnected. Reconnect the switch.
Gone to Sleep
The predetermined sleep time has been exceeded due to inactivity by the user.
Charging
This indication is present when the battery charger is connected. There may
also be an error in the control module. Contact Sunrise Technical Service for
assistance.
1.
In this example, one of the right motor
being disconnected (
fi
g.5.3.1). The hand control
displays the fault and the power chair will not
operate.
2.
To troubleshoot this problem, refer to Table
1, Error Codes.
3.
From this error we see that the problem is
being recorded in the PM (Control Module).
4.
The hand control tells us that motor 2 (M2)
has the error and it is with the brake circuit. Motor
2 is the right motor.
5.
Table 1 informs us to, “Verify that the
right motor is connected to the control module.
Complete the Motor and Gearbox Inspection
section of this manual.”
6.
The next step is to complete the Motor and
Gearbox Inspection section of this manual and
Example of R-10 Fault Isolation
fig. 5.3.1
Содержание Groove
Страница 1: ......
Страница 2: ...2 2006 Sunrise Medical ...
Страница 14: ...14 2006 Sunrise Medical INTRODUCTION ...
Страница 119: ...2009 Quickie Rhythm Service Manual Supplement 2009 Sunrise Medical Inc xxxxxx Rev A PG DRIVES TECHNOLOGY R ...
Страница 132: ...RHYTHM SERVICE MANUAL PAGE 0 14 2009 SUNRISE MEDICAL ...
Страница 138: ...RHYTHM SERVICE MANUAL 2009 PAGE 2 4 SUNRISE MEDICAL Section 2 ...
Страница 144: ...RHYTHM SERVICE MANUAL 2009 PAGE 3 6 SUNRISE MEDICAL ...
Страница 148: ...RHYTHM SERVICE MANUAL 2009 PAGE 4 4 SUNRISE MEDICAL ...
Страница 162: ...RHYTHM SERVICE MANUAL 2009 PAGE 6 6 SUNRISE MEDICAL ...
Страница 168: ...RHYTHM SERVICE MANUAL 2009 PAGE 8 4 SUNRISE MEDICAL Tilt to Interface Mounting Section 8 ...
Страница 179: ...RHYTHM SERVICE MANUAL PAGE 9 1 2009 SUNRISE MEDICAL Section 9 OBP Programming Quick Guide ...
Страница 180: ...RHYTHM SERVICE MANUAL 2009 PAGE 9 2 SUNRISE MEDICAL Section 9 OBP Programming Quick Guide ...
Страница 181: ......
Страница 182: ...182 2006 Sunrise Medical ...