RHYTHM SERVICE MANUAL
PAGE 6.5
2009
SUNRISE MEDICAL
Section 6
Power Shear Does not Operate
1. To test the Shear actuator resistance, locate the
ISM mounted on the back of the wheelchair.
2.
Disconnect the 2 pin connector at actuator
port 1 (
fi
g. 6.5.1)
3.
Check for resistance at this harness,
normal resistance would be 6 to 8 ohms when the
actuator
is in mid-stroke and greater than 1 m ohms when
at
either extremity.
If you have veri
fi
ed resistance at this connection,
Replace the ISM.
Left
Lights
Inhibit
4
Inhibit
5
Right
Lights
1
2
3
4
5
6
Actuator
Channels
Actuator
Channels
4.
If your reading is open, re-test the resistance
at the 6 pin connection leading directly from the
shear actuator (
fi
g 6.5.2) at pins 1 and 5.
If your reading is open, replace the shear
actuator.
If resistance is veri
fi
ed, replace the harness
leading from the 6 pin actuator harness to the
ISM.
fig. 6.5.1
fig. 6.5.2
1
2
3
4
5
6
Содержание Groove
Страница 1: ......
Страница 2: ...2 2006 Sunrise Medical ...
Страница 14: ...14 2006 Sunrise Medical INTRODUCTION ...
Страница 119: ...2009 Quickie Rhythm Service Manual Supplement 2009 Sunrise Medical Inc xxxxxx Rev A PG DRIVES TECHNOLOGY R ...
Страница 132: ...RHYTHM SERVICE MANUAL PAGE 0 14 2009 SUNRISE MEDICAL ...
Страница 138: ...RHYTHM SERVICE MANUAL 2009 PAGE 2 4 SUNRISE MEDICAL Section 2 ...
Страница 144: ...RHYTHM SERVICE MANUAL 2009 PAGE 3 6 SUNRISE MEDICAL ...
Страница 148: ...RHYTHM SERVICE MANUAL 2009 PAGE 4 4 SUNRISE MEDICAL ...
Страница 162: ...RHYTHM SERVICE MANUAL 2009 PAGE 6 6 SUNRISE MEDICAL ...
Страница 168: ...RHYTHM SERVICE MANUAL 2009 PAGE 8 4 SUNRISE MEDICAL Tilt to Interface Mounting Section 8 ...
Страница 179: ...RHYTHM SERVICE MANUAL PAGE 9 1 2009 SUNRISE MEDICAL Section 9 OBP Programming Quick Guide ...
Страница 180: ...RHYTHM SERVICE MANUAL 2009 PAGE 9 2 SUNRISE MEDICAL Section 9 OBP Programming Quick Guide ...
Страница 181: ......
Страница 182: ...182 2006 Sunrise Medical ...