117
©2006 Sunrise Medical
SEATING SECTION
In order to eliminate the re-home switch (previously
found on the Smart Seat), a separate Counter Reset
Switch (CRS) is installed that allows the system to “re-
home” during normal operation. This switch is located
near the bottom of the shear actuator (Fig 3.130).
Complete the following steps to remove/replace or
adjust the Shear CRS.
The Shear CRS may be adjusted by sliding the
1.
switch bracket side to side.
In order for the CRS to work correctly, the roller
2.
switch must contact the button head screw on the
shear actuator mount.
The CRS can be adjusted by loosening the two
3.
screws behind the switch bracket with a 9/64” hex
wrench.
To remove the bracket, remove the two screws
4.
that were used for adjustment.
Follow the wires up to their mating connectors and
5.
disconnect them.
Reverse the procedure to install the new Shear
6.
CRS. The part number is 103952 (Fig 3.131).
Fig 3.130
Fig 3.131
Shear CRS Replacement and Adjustment
Содержание Groove
Страница 1: ......
Страница 2: ...2 2006 Sunrise Medical ...
Страница 14: ...14 2006 Sunrise Medical INTRODUCTION ...
Страница 119: ...2009 Quickie Rhythm Service Manual Supplement 2009 Sunrise Medical Inc xxxxxx Rev A PG DRIVES TECHNOLOGY R ...
Страница 132: ...RHYTHM SERVICE MANUAL PAGE 0 14 2009 SUNRISE MEDICAL ...
Страница 138: ...RHYTHM SERVICE MANUAL 2009 PAGE 2 4 SUNRISE MEDICAL Section 2 ...
Страница 144: ...RHYTHM SERVICE MANUAL 2009 PAGE 3 6 SUNRISE MEDICAL ...
Страница 148: ...RHYTHM SERVICE MANUAL 2009 PAGE 4 4 SUNRISE MEDICAL ...
Страница 162: ...RHYTHM SERVICE MANUAL 2009 PAGE 6 6 SUNRISE MEDICAL ...
Страница 168: ...RHYTHM SERVICE MANUAL 2009 PAGE 8 4 SUNRISE MEDICAL Tilt to Interface Mounting Section 8 ...
Страница 179: ...RHYTHM SERVICE MANUAL PAGE 9 1 2009 SUNRISE MEDICAL Section 9 OBP Programming Quick Guide ...
Страница 180: ...RHYTHM SERVICE MANUAL 2009 PAGE 9 2 SUNRISE MEDICAL Section 9 OBP Programming Quick Guide ...
Страница 181: ......
Страница 182: ...182 2006 Sunrise Medical ...