AQTF-3013-INS-LAB-RevB09
1
AQTF-3013 TENDERFLO
®
WATER CUSHION:
INSTALLATION INSTRUCTIONS
PLEASE SAVE THESE INSTRUCTIONS FOR FUTURE USE
Note: The most current version of these instructions can be found online at
www.grahamfield.com
WARNING: Important! Read and understand these instructions before installing or using the
Lumex
®
Tenderflo
®
Water Cushion. These instructions contain information that will aid you in
achieving maximum therapeutic results. If you do not understand any part of these warnings,
cautions or instructions, contact a healthcare professional for direction in the use of this
product. If the Cushion is not properly installed, personal injury and damage to the Cushion
could result.
WARNING: If components are damaged or missing, contact your dealer immediately. DO NOT
use substitute parts. Use only Lumex replacement parts. The use of non-Lumex replacement
parts could cause personal injury, property damage, and void the warranty.
WARNING: This product has a maximum weight capacity of 250 lb.
WARNING: GF Health Products, Inc. assumes no responsibility for any damage or injury
caused by improper assembly or use of this product.
INTENDED USE
The intended use of the AQTF-3013 is as a support and pressure reduction surface.
INSTALLATION
Setup
Unfold the Cushion and place it on a wheelchair with the fill valve facing up.
Filling
1. Lift up the fill valve cap to open it.
2. Locate a nearby faucet to obtain warm water, approximately 98°F - 99°F (36°C - 37°C).
3. Connect one end of a standard garden hose to the faucet. Screw the male end of the garden hose
into the Cushion's female fill valve.
4. Gently turn on the warm water.
Do not leave the Cushion unattended.
Hold the fill valve upright
until the water level reaches the filling line.
5. Turn off the water when the desired level has been reached.
Do not overfill.
Caution: If you accidentally overfill the Cushion, drain out the excess water immediately.
6. Close the fill valve cap.
Caution: Ensure fill valve cap is securely closed. A loose cap could result in leakage.
7. Mark date filled.
Checking for Proper Fill Level
1. Fill the Cushion with as little water as possible, as previously described. After all the caps are
tightly in place, check for proper fill level:
2. Position the patient on the cushion. Place your hand (palm up) under the Cushion beneath the pa-
tient's buttocks. There should be approximately one to two inches between these two surfaces. The
Cushion should contour to the patient's body and not appear hard from overfilling. If too soft, add
more water, and if too firm, remove water, until appropriate fill is achieved. When properly filled,
position cushion with "
BOTTOM SIDE
" facing downward and arrows facing front of wheelchair.