SEATING SECTION
Additionally there may be encoder errors. Use the
HHP to re-home the chair prior to completing any tests.
Complete the following procedures to test the Shear
Count Reset Switch (CRS).
Locate the QMAC mounted on the back of the
1.
wheelchair (Fig 3.96).
Locate the Shear CRS (B) mounted under the
2.
QMAC (Fig 3.97). Note: this switch is Normally
Open (NO) when the seat-back is fully elevated,
and closes as the seat-back lowers.
3. Verify that the switch is not damaged and free to
3.
operate.
Disconnect the 6-pin female connector (C) leading
4.
to the Shear port of the QMAC(Fig 3.98).
Fig 3.96
Fig 3.97
B
C
Fig 3.98
The back will not drive to the preprogrammed shear value
Содержание Groove
Страница 1: ......
Страница 2: ...2 2006 Sunrise Medical ...
Страница 14: ...14 2006 Sunrise Medical INTRODUCTION ...
Страница 119: ...2009 Quickie Rhythm Service Manual Supplement 2009 Sunrise Medical Inc xxxxxx Rev A PG DRIVES TECHNOLOGY R ...
Страница 132: ...RHYTHM SERVICE MANUAL PAGE 0 14 2009 SUNRISE MEDICAL ...
Страница 138: ...RHYTHM SERVICE MANUAL 2009 PAGE 2 4 SUNRISE MEDICAL Section 2 ...
Страница 144: ...RHYTHM SERVICE MANUAL 2009 PAGE 3 6 SUNRISE MEDICAL ...
Страница 148: ...RHYTHM SERVICE MANUAL 2009 PAGE 4 4 SUNRISE MEDICAL ...
Страница 162: ...RHYTHM SERVICE MANUAL 2009 PAGE 6 6 SUNRISE MEDICAL ...
Страница 168: ...RHYTHM SERVICE MANUAL 2009 PAGE 8 4 SUNRISE MEDICAL Tilt to Interface Mounting Section 8 ...
Страница 179: ...RHYTHM SERVICE MANUAL PAGE 9 1 2009 SUNRISE MEDICAL Section 9 OBP Programming Quick Guide ...
Страница 180: ...RHYTHM SERVICE MANUAL 2009 PAGE 9 2 SUNRISE MEDICAL Section 9 OBP Programming Quick Guide ...
Страница 181: ......
Страница 182: ...182 2006 Sunrise Medical ...