RHYTHM SERVICE MANUAL
2009
PAGE 3.4
SUNRISE MEDICAL
Section 3
VR2 Controller Diagnostics Codes (cont.)
Check that the batteries are fully charged and in good
condition; and check all cables and connections. Check
the connections to the right motor, look for a loose or
damaged connector.
If the reading is short (resistance is less than 10 K
ohms) on any of the readings, proceed to check the
4-pin motor connector. Measure the resistance from
the bottom contact of the red thick wire on the 4-pin
right motor connector to each of the top contacts of the
connectors see (figure 3.4.1). Measure the resistance
from the bottom contact of the black thick wire on the
4-pin right motor connector to each the top contacts of
the connector (below right). If all of the readings are
open, then replace the controller. If any of the readings
are short, then replace the right motor.
Five Bars - Right Motor Wiring Trip
3.4.1
A joystick trip is indicated. Make sure that the joystick is in the center position before switching on the control
system.
Check that the batteries are fully charged and in good condition, examine the joystick for damage. This fault can
be caused by a joystick that fails to center itself due to being dirty, bent or broken. If this is the case, replace the
joystick module.
Note: If replacing the joystick does not resolve the issue, replace the cable connecting the joystick to the
controller.
Seven Bars - Possible Joystick Trip
The Onboard Batteries are being charged with the off-board charger. You will not be able to drive the wheelchair
until the charger is disconnected. You will have to reset the control system by switching off the power and the
Powering up again. The On-Board charger has no indication that the chair is charging, and the chair will not move
until complete.
If the condition still exists after the charger has been disconnected and the chair has been switched off and powered
up again, the Joystick module may be defective.
Six Bars - Charger Connected
Seven Bars + Speed Profile Indicator Communication Error
Inspect wiring between joystick module and controller. Replace the jumper or joystick module with
damaged wiring. If the problem persists replace the controller.
Test 1
Test 2
Test 3
Test 4
Содержание Groove
Страница 1: ......
Страница 2: ...2 2006 Sunrise Medical ...
Страница 14: ...14 2006 Sunrise Medical INTRODUCTION ...
Страница 119: ...2009 Quickie Rhythm Service Manual Supplement 2009 Sunrise Medical Inc xxxxxx Rev A PG DRIVES TECHNOLOGY R ...
Страница 132: ...RHYTHM SERVICE MANUAL PAGE 0 14 2009 SUNRISE MEDICAL ...
Страница 138: ...RHYTHM SERVICE MANUAL 2009 PAGE 2 4 SUNRISE MEDICAL Section 2 ...
Страница 144: ...RHYTHM SERVICE MANUAL 2009 PAGE 3 6 SUNRISE MEDICAL ...
Страница 148: ...RHYTHM SERVICE MANUAL 2009 PAGE 4 4 SUNRISE MEDICAL ...
Страница 162: ...RHYTHM SERVICE MANUAL 2009 PAGE 6 6 SUNRISE MEDICAL ...
Страница 168: ...RHYTHM SERVICE MANUAL 2009 PAGE 8 4 SUNRISE MEDICAL Tilt to Interface Mounting Section 8 ...
Страница 179: ...RHYTHM SERVICE MANUAL PAGE 9 1 2009 SUNRISE MEDICAL Section 9 OBP Programming Quick Guide ...
Страница 180: ...RHYTHM SERVICE MANUAL 2009 PAGE 9 2 SUNRISE MEDICAL Section 9 OBP Programming Quick Guide ...
Страница 181: ......
Страница 182: ...182 2006 Sunrise Medical ...