46
©2006 Sunrise Medical
QUICKIE ELECTRONICS SECTION
Device
Type of input control utilized : hand control,3 Button, SCIM1, SCIM1 or compact joystick.
Forward Speed,
The maximum speed setting for the control module and the minimum setting – to drive the chair in the forward
direction.
Reverse Speed
The maximum speed setting for the control module and the minimum setting – to drive the chair in the reverse
direction.
Turning Speed
The maximum speed setting for the control module and the minimum setting – to drive the chair in the left or right
direction.
Accel Fwd & Rev
How quickly the chair reaches the selected speed up to the maximum. Settings are in 1% increments.
Decel – Fwd & Rev
How quickly the chair comes to a stop from forward or reverse directions. Settings are in 1% increments.
Turn Acceleration
How quickly the chair reaches maximum turning speed, or how quickly the chair responds to a turn command. This
parameter is extremely important as to how sensitive the chair is to input device movements and how easy it is for
the user to control the chair.
Turn Deceleration
How quickly the chair comes to a stop from a turn. Also very important in the control of the chair. Settings are 1%
increments.
Torque Limit
Determines amount of demand or current applied to output of control to motors in response to Drive Input com-
mands.
Menu Item
Description
Range
Drive Pro
fi
les
Determines number of pro
fi
les (QC=1 QR=4)
1-4
Device
Selects Drive input when pro
fi
le is Active
Forward Speed
Determines Fwd Speed settings
5 - 100
Reverse Speed
Determines Rev Speed settings
0 – 100
Turning Speed
Determines Turning Spd settings
0 - 100
Forward Acceleration Determines Fwd Accel Rate
0 - 100
Reverse Acceleration Determines Rev Accel Rate
0 - 100
Forward Deceleration Determines Fwd Decel Rate
0 - 100
Reverse Deceleration Determines Rev Decel Rate
0 - 100
Turn Acceleration
Determines Turn Accel Rate
0 - 100
Turn Deceleration
Determines Turn Decel Rate
0 - 100
Power
Determines allowable current for Motor torque limiting
20 - 100
High Speed Torque
Determines the maximum amount of torque available at the highest speed
20 - 100
Low Speed Torque
Determines the maximum amount of torque available at the lowest speed
20 - 100
Programmable Ranges
Содержание Groove
Страница 1: ......
Страница 2: ...2 2006 Sunrise Medical ...
Страница 14: ...14 2006 Sunrise Medical INTRODUCTION ...
Страница 119: ...2009 Quickie Rhythm Service Manual Supplement 2009 Sunrise Medical Inc xxxxxx Rev A PG DRIVES TECHNOLOGY R ...
Страница 132: ...RHYTHM SERVICE MANUAL PAGE 0 14 2009 SUNRISE MEDICAL ...
Страница 138: ...RHYTHM SERVICE MANUAL 2009 PAGE 2 4 SUNRISE MEDICAL Section 2 ...
Страница 144: ...RHYTHM SERVICE MANUAL 2009 PAGE 3 6 SUNRISE MEDICAL ...
Страница 148: ...RHYTHM SERVICE MANUAL 2009 PAGE 4 4 SUNRISE MEDICAL ...
Страница 162: ...RHYTHM SERVICE MANUAL 2009 PAGE 6 6 SUNRISE MEDICAL ...
Страница 168: ...RHYTHM SERVICE MANUAL 2009 PAGE 8 4 SUNRISE MEDICAL Tilt to Interface Mounting Section 8 ...
Страница 179: ...RHYTHM SERVICE MANUAL PAGE 9 1 2009 SUNRISE MEDICAL Section 9 OBP Programming Quick Guide ...
Страница 180: ...RHYTHM SERVICE MANUAL 2009 PAGE 9 2 SUNRISE MEDICAL Section 9 OBP Programming Quick Guide ...
Страница 181: ......
Страница 182: ...182 2006 Sunrise Medical ...