QIAsymphony SP/AS Consolidated Operating Guide 12/2017
66
4.
Press
Yes
to continue.
The
Eluate Drawer/Elution Slot/Change Rack X
screen is displayed. The
rack on the selected slot is removed.
5.
Remove the elution rack from the elution slot.
If the elution rack was used with an elution adapter, also remove the adapter.
6.
To unload additional elution racks, repeat the process.
7.
When all elution racks have been unloaded, close the “Eluate” drawer.
The
Eluate Drawer/Elution Slot/Configure Rack X
screen appears.
8.
Press
OK
.
The QIAsymphony SP performs an inventory scan of the “Eluate” drawer.
Afterwards the
Sample Preparation/Overview
screen is displayed. .
Important
If eluate cooling in “Elution slot 1” was turned on, it will be turned off as soon as
the
OK
button or
Yes
button is pressed.
8.4
Loading the “Reagents and Consumables” drawer
The “Reagents and Consumables” drawer accommodates all consumables and reagents required
for the protocol run.
Before starting a protocol run, the drawer must be loaded with the appropriate reagents in
prefilled, sealed reagent cartridges, sample prep cartridges, 8-Rod Covers, and disposable filter-
tips. In some cases, an Accessory Trough and buffer bottle may be required.
Содержание QIAsymphony AS
Страница 213: ...QIAsymphony SP AS Consolidated Operating Guide 12 2017 213 This page intentionally left blank...
Страница 214: ...QIAsymphony SP AS Consolidated Operating Guide 12 2017 214 This page intentionally left blank...
Страница 216: ...QIAsymphony SP AS Consolidated Operating Guide 12 2017 216 www qiagen com...