QIAsymphony SP/AS Consolidated Operating Guide 12/2017
146
Press
Run
in the assay setup
Overview
screen.
If an inventory scan was performed after pressing the
Queue
button, provided that the validation
showed no error and nothing was changed after that point, the inventory scan will be skipped
and the assay run starts immediately.
If an inventory scan was not performed after pressing the
Queue
button, a message will appear
asking whether an inventory scan should be performed for each drawer.
See Section 12.4.7 for detailed information about validating the assay run.
12.4.11
Removing assays after an independent run
When an assay run is completed or canceled, the assays must be removed from the “Assays”
drawer. The assays will not be automatically removed from the QIAsymphony AS.
If the status of a run is shown as
QUEUED
,
STOPPED
, or
COMPLETED
, the assay rack(s) and
adapter(s) can be removed.
It is possible to remove assays after an independent run in the same manner that they are
removed after an AS run; see Section 12.3.5. Alternatively, follow the steps below.
1.
Open the “Assays” drawer.
The
Assay Setup/Loading Information screen
appears.
Содержание QIAsymphony AS
Страница 213: ...QIAsymphony SP AS Consolidated Operating Guide 12 2017 213 This page intentionally left blank...
Страница 214: ...QIAsymphony SP AS Consolidated Operating Guide 12 2017 214 This page intentionally left blank...
Страница 216: ...QIAsymphony SP AS Consolidated Operating Guide 12 2017 216 www qiagen com...