QIAsymphony SP/AS Consolidated Operating Guide 12/2017
112
Modifying an integrated run
If an integrated run has already been defined, the
Integrated Run
screen displays the status of all
defined integrated batches and the relationship between SP and AS batches.
1.
Press
Modify Run
. The
Integrated Setup
screen appears and
displays an overview of the defined batches.
2.
Use the
Remove AS Batch
button to remove an AS batch from the
integrated run of the related SP batch.
This button is unavailable if the AS batch has been started.
3.
The
Edit Samples
button allows bar code reading errors for
sample tubes to be resolved. In addition, sample IDs, sample
types, and sample labware can be modified.
4.
Assign the assay to all samples of a batch via the favorite buttons.
5.
Define assays for a batch.
6.
Use the
Create AS Batch
button to assign an AS batch to one or
more SP batches.
Important
It is possible to change the order in which batches of an integrated run are
processed by manually unloading, reloading, and redefining an integrated
batch.