QIAsymphony SP/AS Consolidated Operating Guide 12/2017
109
After modifying a parameter, the hand symbol is displayed in the active assay tab. After
modifying assay parameters, a hand symbol appears.
6.
If the parameters for more than one assay need to be modified, repeat steps 2–5 for the other
assay(s).
7.
To overwrite the default elution volume, select the required elution volume
from the
Eluate Volume
list on the left side of the screen by pressing the
appropriate button.
8.
Press
OK
to save the changes and to return to the
Assay Assignment
screen.
Important
If the parameters are modified, the changes will not be saved in the assays. They
will be used for the current run only. To change parameters in an assay for future
runs, use the
Process Definitio
n editor tool of the QIAsymphony Management
Console.
Assigning assays using work lists
If a work list(s) is used, assays are automatically assigned to samples via their written bar code(s),
as defined in the work list(s). Those batches that have assays assigned to them by work lists are
Содержание QIAsymphony AS
Страница 213: ...QIAsymphony SP AS Consolidated Operating Guide 12 2017 213 This page intentionally left blank...
Страница 214: ...QIAsymphony SP AS Consolidated Operating Guide 12 2017 214 This page intentionally left blank...
Страница 216: ...QIAsymphony SP AS Consolidated Operating Guide 12 2017 216 www qiagen com...