QIAsymphony SP/AS Consolidated Operating Guide 12/2017
64
Adapter placed onto the transfer frame on slot 1 of the “Eluate” drawer.
The “Eluate” drawer is locked during:
Transfer of eluates from sample prep cartridges to the elution rack
During the inventory scan of the “Eluate” drawer
During transfer of eluates from the QIAsymphony SP to the QIAsymphony AS via the transfer
module
During an integrated run
At all other times, the “Eluate” drawer can be opened or closed.
8.3.4
Unloading the “Eluate” drawer
Elution racks must be manually unloaded from the “Eluate” drawer.
If using QIAsymphony SP/AS instruments in integrated run mode, an elution rack on “Elution
slot 1” will be automatically transferred from the QIAsymphony SP to the AS module to start the
reaction setup. Afterwards, the elution rack will be automatically transferred back to the
QIAsymphony SP “Eluate” drawer.
If using QIAsymphony SP/AS instruments in the independent mode, the elution rack can be
directly transferred to the “Eluate and Reagents” drawer of the QIAsymphony AS by using the
Transfer
button.
To transfer elution racks from any elution slot other than “Elution slot 1”, a manual transfer must
be performed. For increased flexibility, elution racks can be removed from the “Eluate” drawer
before a protocol run in independent mode has finished. As soon as eluates have been
transferred to an elution rack, the elution rack can be removed from the drawer.
Содержание QIAsymphony AS
Страница 213: ...QIAsymphony SP AS Consolidated Operating Guide 12 2017 213 This page intentionally left blank...
Страница 214: ...QIAsymphony SP AS Consolidated Operating Guide 12 2017 214 This page intentionally left blank...
Страница 216: ...QIAsymphony SP AS Consolidated Operating Guide 12 2017 216 www qiagen com...