QIAsymphony SP/AS Consolidated Operating Guide 12/2017
131
Assigning a rack file
1.
Press a “Sample” slot to select it. Ensure that only one “Sample” slot is selected. A selected
“Sample” slot is highlighted dark yellow.
2.
To deselect a “Sample” slot, press it. It will then appear pale yellow.
3.
Press
Rack Files
.
The
Select rack file
list appears.
4.
Press a rack file to select it from the list.
There are 2 types of rack files —
Sample Racks
and
Assay Racks
.
Sample Racks
are standard
sample rack files for defining an assay run. In some cases, the assay rack can be used as a
sample rack (e.g., for setting up two-step RT-PCR assays). In this case, an
Assay Rack
can be
selected.
5.
When the assay rack file is selected, an information message appears.
Press
Yes
to continue.
The selected rack file will be assigned to the selected “Sample” slot. The rack type and rack ID
that are defined in the selected rack file will be assigned to the selected “Sample” slot. The
“Sample” slot will now appear blue and the
Next
button becomes active.