QIAsymphony SP/AS Consolidated Operating Guide 12/2017
188
14.5.3
Cleaning the QIAsymphony SP/AS hoods
To clean the hoods of QIAsymphony SP/AS instruments, wipe the surface with a soft lint-free cloth
moistened with deionized water, or use wipes soaked with Mikrozid Sensitive Liquid. Then wipe
dry with a dry soft lint-free cloth or paper towel.
Important
Do not use ethanol-based disinfectant; use distilled water or Mikrozid Sensitive
Liquid only.
14.5.4
Cleaning the tube carriers (SP)
1.
Remove tube carriers, adapters, and inserts and soak them in disinfectant (e.g., Gigasept
Instru AF). Incubate for at least 15 min, then rinse with water and dry with paper towels.
2.
Check the condition of the bar code labels and ensure that they are not scratched.
14.5.5
Cleaning the optical sensor (SP)
Wipe the window of the optical sensor with a lint-free cloth. Moisten the cloth with 70% ethanol if
required.
14.5.6
Magnetic head (SP)
1.
Remove the cover from the magnetic head.
2.
Move the magnetic head up and carefully push the rod cover holder down.
Содержание QIAsymphony AS
Страница 213: ...QIAsymphony SP AS Consolidated Operating Guide 12 2017 213 This page intentionally left blank...
Страница 214: ...QIAsymphony SP AS Consolidated Operating Guide 12 2017 214 This page intentionally left blank...
Страница 216: ...QIAsymphony SP AS Consolidated Operating Guide 12 2017 216 www qiagen com...