QIAsymphony SP/AS Consolidated Operating Guide 12/2017
121
9.
Press
Load
The
Assay Setup/Loading information
screen appears again.
The loaded slot is now shown in blue.
See procedures “Entering reagent kit bar codes” and “
Defining customized kit bar codes
” below.
Entering reagent kit bar codes
To enter a reagent kit bar code for each assay, follow the steps below.
1.
Switch to
List View
, or press the
Scan Kit Bar Code
button.
2.
Press the appropriate tab to select an assay.
3.
Press the
Kit bar code
field.
4.
Manually enter the bar code, or enter a bar code using the bar code scanner.
5.
Press
OK
to return to the
Loading Reagents
screen. If the bar code scanner was used, the
Loading Reagents
screen will automatically reappear.
6.
The software validates the kit bar code of known format and checks the lot number and
expiration date.
Important
Multiple kit bar codes for one assay must be separated by a semi colon. In this
case, validation of lot number and expiration date will not be performed.
Important
Entered kit bar codes, including additional information (i.e., expiration date,
product number, and lot number), are tracked in the result file.
Important
If the entered kit bar code does not follow a recognized format, a message will
appear, asking whether to accept the bar code. Press
OK
to continue.
Содержание QIAsymphony AS
Страница 213: ...QIAsymphony SP AS Consolidated Operating Guide 12 2017 213 This page intentionally left blank...
Страница 214: ...QIAsymphony SP AS Consolidated Operating Guide 12 2017 214 This page intentionally left blank...
Страница 216: ...QIAsymphony SP AS Consolidated Operating Guide 12 2017 216 www qiagen com...