QIAsymphony SP/AS Consolidated Operating Guide 12/2017
162
Error code
Description
Comments and suggestions
70505
Some of the selected combinations of
ACS and elution rack types that you
selected are marked as “not
recommended” [List of not
recommended combinations].
Press
No
to stay in this dialog and
change the elution rack or continue
without any changes by pressing
Yes
.
Perform one of the following:
Press
Yes
to ignore the message.
Press
No
and change the elution rack.
Use the
Labware Browser
menu or the labware lists to identify
which labware is recommended for use with which protocols.
Labware lists are available for download at
www.qiagen.com/goto/QIAsymphony.
The “Supervisor” can go to the
Configuration
menu and
disable the parameter
Check combination of protocol and
recommended labware during run definition?
in the
Process
SP 2
tab.
71840
Leaving the dialog without successful
inventory scan will lead to
unexpected results. Do you really
want to skip the scan and cancel the
dialog?
If you do not perform an inventory scan of the “Eluate”
drawer, eluates may be transferred to a slot that does not
contain an elution rack. Run an inventory scan of the “Eluate”
drawer to make sure that the selected elution slots contain
elution racks.
Note
: If the inventory scan of the “Eluate” drawer persistently
fails, contact QIAGEN Technical Services.
72002
A rack file is already assigned to a
sample slot.
It is not possible to change the information for this slot as a
rack file is assigned. To modify information for this slot, restart
the assay definition process without a rack file.
72202
Errors have been identified in the
work lists shown below. Please
ensure that work lists contain a
unique elution rack ID per batch and
an elution rack ID for all samples.
Work lists with inconsistencies or
errors: Worklist_X (missing required
elution rack ID for samples: “Y”;
multiple elution rack IDs: “Z”.
A work list file defines 2 different required elution rack IDs for
the samples of the batch currently being defined. This work list
therefore cannot be used.
Either change the required elution rack ID in the work list or
define the batch in that way that the samples to be processed
uses just one elution rack ID.
72204
The following samples in work list X
are not present: Y. Please place the
missing samples on the sample input
or correct the work list.
A sample was listed in a work list file, but this sample is not
part of the batch being defined. The work list file can therefore
not be used.
Perform one of the following:
Add the missing sample(s) to the batch.
Delete the missing sample(s) from the work list.
Set the configuration parameter
Allow partial use of work
lists
or
Allow processing of samples without a work list
entry
to
Yes
in the
Process SP 3
tab of the
Configuration
menu. This can only be performed by the “Supervisor”.
72208
The following samples cannot be
processed because they are not listed
in any work list: X. Please remove
these samples from the sample input
or correct the work list.
A sample that is part of the batch currently being defined is
not listed in a work list file. The batch can therefore not be
defined.
Perform one of the following:
Remove the sample from the batch.
Add the sample to a work list.
Set the configuration parameter
Allow processing of
samples without a work list entry?
to
Yes
in the
Configuration
menu. This can only be performed by the
“Supervisor”.
75108
Unable to assign one rack to more
than one slot. A rack ID may have
been used more than once.
It is not possible to assign the same rack file to more than one
“Sample” slot.
Deselect one “Sample” slot and repeat the assignment of the
rack file.