QIAgility Software
QIAgility User Manual 11/2011
5-117
Use selected
wells INSIDE
range
Instructs the software to select rows where
the concentration value is within the
defined concentration range.
Use selected
wells OUTSIDE
range
Instructs the software to select rows where
the concentration value is outside the
defined concentration range.
The second filter is based on membership in at least one
sample bank selected from the file. This function,
“Members of selected bank columns only”, is available
when “Bank from column” or “Banks named from
membership columns” is selected. When the filter is
activated, the “Import Preview” pane shows only rows
with a non-blank value in the selected “Bank from
column” or with a non-blank value in any one of the
columns listed at “Banks named from membership
columns”.
10.
The “Import Preview” panel displays the samples to be
imported. Click the “Import” button to import the data.
The imported data will be displayed in the right-hand
pane.
Note
: No data is imported from a row in which all fields
are empty.
Note
: Existing data will be overwritten by any imported
data that is located in the same well.
Note
: Multiple sample banks
simultaneously
imported
using the “Banks named from membership columns”
function. Alternate strategies allow multiple sample
banks to be
successively
populated from the same
import file. For example this can be done using the
“Bank from column” function, where several columns,
one per sample bank, are present in the file and each
row includes the bank name in the corresponding
column to indicate its membership in that bank. Set the
“Bank from column” and activate the “Members of
selected bank columns only” filter. Click on the “Import”
button. Repeat for each sample bank column in the file.
It can also be done using the concentration filter
function. Arrange the import file so that it has a column
Содержание QIAgility
Страница 10: ...Contents Contents 8 QIAgility User Manual 11 2011 This page intentionally left blank...
Страница 20: ...Safety Information 1 10 QIAgility User Manual 11 2011 This page intentionally left blank...
Страница 40: ...General Description 3 16 QIAgility User Manual 11 2011 This page intentionally left blank...
Страница 62: ...Installation Procedures 4 22 QIAgility User Manual 11 2011 This page intentionally left blank...
Страница 211: ...QIAgility Software QIAgility User Manual 11 2011 5 149...
Страница 212: ...QIAgility Software 5 150 QIAgility User Manual 11 2011 This page intentionally left blank...
Страница 254: ...Instrument Setup and Calibration 7 20 QIAgility User Manual 11 2011 This page intentionally left blank...
Страница 260: ...Maintenance Procedures 8 6 QIAgility User Manual 11 2011 This page intentionally left blank...
Страница 274: ...Troubleshooting 9 14 QIAgility User Manual 11 2011 This page intentionally left blank...
Страница 278: ...Glossary 10 4 QIAgility User Manual 11 2011 This page intentionally left blank...
Страница 284: ...Appendix A A 6 QIAgility User Manual 11 2011 This page intentionally left blank...
Страница 286: ...Appendix B B 2 QIAgility User Manual 11 2011 This page intentionally left blank...
Страница 298: ...Appendix C C 12 QIAgility User Manual 11 2011 This page intentionally left blank...
Страница 326: ...Appendix G G 4 QIAgility User Manual 11 2011 This page intentionally left blank...
Страница 328: ...Appendix H H 2 QIAgility User Manual 11 2011 This page intentionally left blank...
Страница 332: ...Index Index 4 QIAgility User Manual 11 2011 This page intentionally left blank...