EZ1 Advanced XL User Manual 11/2017
85
7
Troubleshooting
7.1
EZ1 Advanced XL door errors
If the EZ1 Advanced XL door is open when you start a protocol run, the following error screen will
be displayed.
ERROR: Door open!
Close the door and
retry.
ESC: Retry
Close the door and press
ESC
to continue with the protocol run.
If the door sensor indicates that the EZ1 Advanced XL door is closed but the locking mechanism is
defective, the following error screen will be displayed.
ERROR: Can not lock.
Close the door and
retry.
ESC: Retry
Press
ESC
to retry. If the error persists, contact QIAGEN Technical Services.
7.2
Other errors
If a protocol run is interrupted due to an error:
The red LED flashes
An alarm sounds
The display will show an error message
ERROR: ErrCode
Line: LineNo
ESC: Next
The first line shows the Error Code. Refer to the error code list in Section 7.3 for more information
about the error.
The second line indicates the line number of the protocol at which the error occurred.