EZ1 Advanced XL User Manual 11/2017
86
Record the error code and the line number, and contact QIAGEN Technical Services. Then reset
the EZ1 Advanced XL by following the procedure given below.
1.
Press
ESC
to display the main menu.
2.
Ensure that the EZ1 Advanced XL door is closed.
3.
Press
2
to select the manual function.
01 Apr 2017 10:30
START: Run
1: UV
2: Man
3: Test
4: Setup
4.
Press
2
to return the tips to the tip rack and to return the modules to their home positions.
MANUAL OPERATION
1: Home
2: Return Tip
3: Clean
4: Resend
ESC: Back
5.
Press
ESC
to return to the main screen.
Another protocol can now be run.
Note
: It is not possible to continue a protocol run that has been interrupted due to an error.
7.3
Error codes
Error code
Description
10
Home sensor not activated during an operation other than a protocol.
11
End-limit sensor not activated during an operation other than a protocol.
12
Home sensor of the pipettor head (Z-axis) not activated during a protocol run.
13
Home sensor of the plunger unit (P-axis) not activated during a protocol run.
14
Home sensor of the magnet unit (M-axis) not activated during a protocol run.
15
Home sensor of the worktable (Y-axis) not activated during a protocol run (e.g.,
if the knurled knob used for transportation was not removed, this error will occur
[see Section 4.3.1]).
16
End-limit sensor of the pipettor head (Z-axis) not activated during a protocol run.
17
–
18
–
19
End-limit sensor of the worktable (Y-axis) not activated during a protocol run.