EZ1 Advanced XL User Manual 11/2017
56
Make sure that the tips, tip holders and tip rack are correctly positioned on the worktable.
5.7
Using the UV lamps
At the end of a protocol run, a display message will give you the option of starting a UV
decontamination run. Alternatively, you can start a UV decontamination run manually by switching
on the UV lamp as described below.
Note
: UV decontamination helps to reduce possible pathogen contamination of the EZ1 Advanced
XL worktable surfaces. The efficiency of inactivation has to be determined for each specific organism
and depends, for example, on sample layer thickness and sample type. QIAGEN cannot guarantee
complete eradication of specific pathogens.
5.7.1
Switching on the UV lamps
1.
Make sure that the EZ1 Advanced XL door is closed.
01 Apr 2017 10:30
START: Run
1: UV
2: Man
3: Test
4: Setup