EZ1 Advanced XL User Manual 11/2017
64
5.9.2
Test heating block
This function tests if the heating block of the EZ1 Advanced XL is operating.
Press
2
in the test screen to choose the “Temp” operation.
TEST
1: Axis
2: Temp
3: Serial
4: Version
ESC: Back
The following screen appears.
TEST: TEMPERATURE
Set temp: SS.S C
Up, Dn: Set temp.
START: Run ESC: Back
SS.S
indicates the set temperature in degrees Celsius. Press the up arrow or the down arrow to
increase or decrease the temperature. The upper limit is 99°C. Press
START
to start the heating
process.
The following screen appears.
TEST: TEMPERATURE
Temp: SS.S C
Actual: RR.R C S
ESC: Back
SS.S
indicates the set temperature, and
RR.R
indicates the current temperature.
S
indicates the result,
showing
O
if the temperature is within a given range or
X
if the temperature is outside the given
range.
5.9.3
Test serial port
Press
3
in the test screen to choose the “Serial” operation.
TEST
1: Axis
2: Temp
3: Serial
4: Version
ESC: Back
The serial port test screen appears.
TEST: SERIAL PORT
1: PC/Printer
2: Bar code
ESC: Back