EZ1 Advanced XL User Manual 11/2017
62
5.9
Test function
Press
3
in the main menu to select the test function.
01 Apr 2017 10:30
START: Run
1: UV
2: Man
3: Test
4: Setup
The test screen appears.
TEST
1: Axis
2: Temp
3: Serial
4: Version
ESC: Back
5.9.1
Test axis
Press
1
in the test screen to choose the “Axis” operation.
TEST
1: Axis
2: Temp
3: Serial
4: Version
ESC: Back
The following test procedures will be carried out by the EZ1 Advanced XL.
Move modules to their home positions.
Pick up tip 2 and release it.
Pick up tip 1 and move it to sample position and Proc 1 position.
Move the magnet unit to “narrow” position and then return to home.
Aspirate and dispense to test the pipetting unit.
Move to Proc 2–Proc 11 positions (see figure below).
Move to elution position.
Release tip 1.