Page 49
If the interval timer is activated, the current setting for the
interval time and interval counts can be seen. To change
these settings, press on the respective row.
The interval time can be set between 5 and 1000
seconds.
The interval counts can be set between 2 and 500.
13.5 Jobs
Press
and
and select
„Jobs“ to relate
measurements to a specific job.
Depending on the type of your organization (contractor,
road laboratory, manufacturer etc.) a job could be a
specific client, the development of product “xy”, a special
project and so on.
If you do not change anything in this menu, the job
“default” will be indicated in the measuring mode.
To delete a job,
press on the row „Delete job“ and select
the job to be deleted. It is recommended to make a back-
up copy on your computer or on a USB flash drive.
Confirm the message if you really want to delete the job.
13.6 Setup
13.6.1 User