Page 19
Label of the calibration standard with following
information:
-
article number
-
serial number
-
illumination (entrance) angle
(β)
-
observation angles
-
calibration values
-
expiry date of calibrationstandard
calibration standard
– measured in factory
After the expiry date of the calibration standard
a factory calibration of the ZRS 6060 and its corresponding calibration
standard is required. Contact either
Proceq
or your authorized Proceq agent.
The calibration standard delivered with the ZRS 6060 is not interchangeable
and is valid only for the delivered instrument.
Always protect the calibration standard from dust, moisture and other
environmental factors, keep it mounted on the ZRS 6060 and store the
instrument in its shoulder bag.
7.2 Cleaning of the calibration standard
The calibration standard can be cleaned using standard window cleaner and a soft
cloth.
A damaged or a polluted calibration standard may cause incorrect calibration
and therefore incorrect measuring results.
7.3 Calibration on calibration standard
Should the unit indicate that the calibration is outdated
or it is required by the regulations, the ZRS 6060 has
to be calibrated. The calibration interval can be
changed as described in chapter 13.6.9
on page 53.
Содержание ZRS 6060
Страница 1: ...ZRS 6060 Retroreflectometer Firmware as from v 1 0 Instruction manual...
Страница 5: ...Page 5 16 3 Adjustment of display inclination resistance 61 17 Technical specifications 61 Glossary 63...
Страница 13: ...Page 13 4 Device overview 4 5 1 11 3 10 7 6 14 9 8 2 13 12 15 2 2 4 5 5 5 6 6 6 7 8 9 10 14 16...
Страница 45: ...Page 45...
Страница 46: ...Page 46 13 Menu 13 1 Menu structure...
Страница 67: ......