Page 23
Here the value of the second standard can be inserted.
This might be a known reference value of an internal
sample or a value measured by an official institute. After
the input has been confirmed by tapping
„OK“, the menu
item „calibration settings“ will be displayed automatically.
Please observe that the R
A
value of samples may
change over time as retroreflective materials are
subject disruption. In this case you either need to
re-certify the value of the second calibration
standard on a regular basis or replace it.
The appropriate name of the second calibration standard
can be entered here. Afte
rwards confirm by „OK“. After
confirmation the menu item „calibration settings“ will be
displayed automatically. Tap the backward arrow
to
get back to the calibration function.
After the values for every observation angle have been
set accordingly, you are able to carry out a calibration on
your second standard in the same way as calibrating on
the regular calibration standard.