Page 66
11 Error messages
Error
Description
Solution
6100
Carbox fan error
Fan in power
supply unit is
blocked
Make sure the fan is not blocked
mechanically.
6110
Lamp fan blocked
Halogen lamp
fan is blocked
Make sure the fan is not blocked
mechanically. Remove cooling air inlet
and check visually if fan is blocked.
6112
Lamp voltage error
Lamp voltage is
out of range
Make sure the measuring head is plugged
in correctly. Clean contacts on connector
plug with contact spray.
6113
Lamp power error
Lamp power is
out of range
Make sure the lamp is not broken. Check
visually if the lamp is on. If not it might be
broken.
6114
Chopper speed error
Chopper speed
is out of range
Make sure the chopper is not blocked.
Restart the system.
6116
CRio fan blocked
CRio vent is
blocked
Make sure the fan is not blocked
mechanically. Restart the system.
7110
Measuring head not
connected
Unable to
connect to the
measuring head
Make sure the current NetSetMan profile
is set to ZDR 6020. Restart the system.
Check the LEDs at the network plug of the
docking station.
7210
LAN connection not
available
Ethernet
Controller not
active or no
power on
docking station
Check your network adapter and the
docking station’s power cable.
7310
Data communication
error
Data format
error
Restart the system.
7312
Sensor saturation
Too much light
on sensor
Adjust alignment of measuring head.
7420, 7422
State machine error
Cannot
measure
because lamp,
chopper or
FIFO is not OK
Restart the system.
7501
GPS NL-404P serial
port error
Cannot open
GPS serial port
Select the right COM port in the app
settings.
7502
GPS Trimble serial
port error
Cannot open
GPS serial port.
Select the right COM port in the app
settings.
7511
GPS NL-404P error
No GPS Data
available
Make sure the GPS sensor is connected
to the docking station.
7515
GPS Trimble error
No GPS data
available
Make sure the GPS antenna is plugged in
correctly in the power supply unit (car