Page 61
The lamp inside the measuring head may be replaced by a technician. Please
follow the instructions below:
Tools and parts needed:
1 Allen key 3 mm
1 Allen key 2.5 mm
1 Allen key 2 mm
1 Spare lamp
Note: The spare lamps are aligned optically and
then glued to the lamp holder part. This
alignment process may only be performed at
Proceq SA Factory.
Open the measuring head housing:
1.
Loosen the three screws on the right side of
the cable on the back side (Allen key 2
mm).
2.
Loosen the three central screws on the
front side (Allen key 3mm).
3.
Loosen the four screws on the cooling air
outlet on the front side (Allen key 2.5mm).