ALARMS, SAFETY, AND SELF-DIAGNOSIS
Tables 6–8 show the behaviour of the alarm (buzzer) and ring back function when pressing the BUZZER
STOP key.
Table 6. In cases other than those covered in Tables 7 and 8
Remote Alarm setting
Ring Back
setting
Buzzer from CO
2
incubator
Remote Alarm
When pressing
the BUZZER
STOP key
When the Ring Back
set time passes
When pressing
the BUZZER
STOP key
When the Ring Back
set time passes
ON: Remote alarm setting not linked with
BUZZER STOP key
ON
OFF
(Alarm is not
cancelled)
ON
ON
ON
(Under continuation)
OFF OFF
OFF: Remote alarm setting linked with
BUZZER STOP key
ON ON
OFF (Alarm is
not cancelled)
ON
OFF OFF
OFF
Note:
Resolve the cause of the alarm, as the alarm itself is not deactivated by pressing the
BUZZER STOP
key.
Table 7. In the case of high limit temperature alarm or Err38
Remote Alarm setting
Ring Back
setting
Buzzer from CO
2
incubator
Remote Alarm
When pressing
the BUZZER
STOP key
When the Ring Back
set time passes
When pressing
the BUZZER
STOP key
When the Ring Back
set time passes
ON: Remote alarm setting not linked with
BUZZER STOP key
ON
ON
ON
(Under continuation)
ON
(Continue)
ON
(Under continuation)
OFF
OFF: Remote alarm setting linked with
BUZZER STOP key
ON
OFF
Note:
Close the outer door when Err38 is activated.
Table 8. In cases of Err01, Err02, Err11, Err12, Err18, or door alarm
*1
Remote Alarm setting
Ring Back
setting
Buzzer from CO
2
incubator
Remote Alarm
When pressing
the BUZZER
STOP key
When the Ring Back
set time passes
When pressing
the BUZZER
STOP key
When the Ring Back
set time passes
ON: Remote alarm setting not linked with
BUZZER STOP key
ON
OFF
(Alarm is
cancelled)
OFF
(Alarm is already
cancelled)
OFF
(Alarm is
cancelled
*1
)
OFF
(Alarm is already
cancelled
*1
)
OFF
OFF: Remote alarm setting linked with
BUZZER STOP key
ON
OFF
*1: The remote alarm does not work for the door alarm.
Note:
When Err01 is activated, connect the new CO
2
gas cylinder and press the BUZZER STOP key to
stop the buzzer. In addition, when the optional MCO-50GC is installed and the gas supply is switched to
the reserve gas cylinder, press the BUZZER STOP key and replace the gas cylinder.
ROUTINE MAINTENANCE
To use this unit in a clean condition, clean the chamber and all the inner attachments at least once a month.
1.
Remove all the inner attachments by the procedures shown on pages 23 and 24.
2.
Clean the chamber and all the inner attachments by the procedures shown on page 22.
3.
Install all the inner attachments by the procedures shown on page 25.
●
When there is excessive dirt, contact our sales representative or agent.