8
PRECAUTIONS FOR SAFE OPERATION
If the unit is to be stored unused in an unsupervised area for a long period,
ensure that children do
not have access and that doors cannot be closed completely.
Ask a qualified contractor to carry out disassembly and disposal of the unit.
Leaving the unit in
a location that can be accessed by third parties may result in unexpected accidents (e.g. the unit may
be used for unintended purposes).
Do not leave the plastic bags used for packing in a place where they can be reached by small
children
as this may result in unexpected accidents such as suffocation.
Use the reagent specified by our company for H
2
O
2
decontamination
. Using a different H
2
O
2
solution may result in explosion or damage to the incubator.
When performing H
2
O
2
decontamination,
securely close the internal and external doors
. Failure to
do so may be harmful to health due to leakage of H
2
O
2
gas.
During H
2
O
2
decontamination
, plug the access hole with the silicon cap that is provided.
Failure to
do so may be harmful to health due to leakage of H
2
O
2
gas.
Always use the removal power supply cord that is provided.
Other power supply cord may cause
electric shock or fire.
When using CO
2
/N
2
gas for control,
make sure that there is adequate ventilation
. Using CO
2
/N
2
gas
in a small room without adequate ventilation may cause gas poisoning or oxygen deprivation. In
addition, when opening the incubator doors, do not directly inhale the air in the chamber.
When connecting a gas cylinder to the incubator,
confirm the gas type
.
Confirm that the
connections are secure and that no gas will leak
.
Be sure to use the specified pressure
. Using
an incorrect gas or pressure may result in explosion or fire, or in gas poisoning or oxygen deprivation
due to gas leak.
O
2
gas
increases the susceptibility of substances to burn
. Take care of the handling of flame in a
room where the incubator is installed.
Install the incubator in a location with adequate ventilation.
If adequate ventilation cannot be
provided, then install an alarm system using CO
2
and O
2
densitometers.
Do not look directly at UV light.
UV light is harmful to the eyes.
Never start H
2
O
2
decontamination when “Warning: UV Bulb Life” is displayed in the message display
area. The UV resolve is not sufficient.
Do not use the unlock key to unlock the outer door during H
2
O
2
decontamination or during H
2
O
2
gas resolve by UV.
Doing so may cause harm to health from H
2
O
2
gas leakage.
WARNING