101
PERFORMANCE
Product name
O
2
/CO
2
Incubator MCO-50ML
Model number
MCO-50ML-PE
Temperature control range
Ambient temperature plus 5
o
C to max. 50
o
C
*1
(ambient temperature: 5
o
C to 35
o
C)
Temperature distribution
±0.25
o
C (ambient temperature: 23
o
C, setting: 37
o
C, CO
2
: 5 %, O
2
: 5 %, no load)
Temperature variation
±0.1
o
C (ambient temperature: 23
o
C, setting: 37
o
C, CO
2
: 5 %, O
2
: 5 %, no load)
CO
2
setting range
0 % to 20 %
CO
2
variation
±0.15 % (ambient temperature: 23
o
C, setting: 37
o
C, CO
2
: 5 %, O
2
: 5 %, no load)
O
2
setting range
1 % to 18 % and 22 % to 80 %
O
2
variation
±0.2 % (ambient temperature: 23
o
C, setting: 37
o
C, CO
2
: 5 %, O
2
: 5 %, no load)
Chamber humidity
95 %R.H.±5 %R.H.
Applicable environmental
condition
Temperature: 5
o
C to 35
o
C, Humidity: 80 %R.H. max.
(The optimum performance may not be obtained
if the ambient temperature is not above 15
o
C)
Noise level
29 dB (A scale)
Power consumption
Max. 210 W
Heat emission
Max. 650 kJ/h
Rated voltage, frequency
AC 220 V–240 V, 50 Hz/60 Hz
Amperage
Max. 1.0
A
*1: When set temperature is 37
o
C, ambient temperature must be 32
o
C or less. Regardless of ambient
temperature, the maximum value of the temperature control range is always 50
o
C.
Notes:
The unit with CE mark complies with EU directives.
Based on our measuring method.
Default calibration conditions: 37
o
C, CO
2
: 5 %, O
2
: 5 %
When using under other conditions, we recommend calibration under the conditions of use.
We recommend calibration every year.