BASIC PARAMETERS
Setting O
2
gas
The O
2
gas control mode during the normal operation of the incubator can be set as follows. The incubator
automatically starts operation using these settings after power-on.
Setting values: ON or OFF
Factory setting: OFF
1. On the home screen, press the menu key (MENU/HOME).
▶
The left side of the display changes to the Menu screen.
2. When the cursor on the menu screen is on Set Value [Fig. 1], press
the enter key (ENTER).
▶
The display changes to the Set Value screen.
3. On the Set Value screen, move the cursor to O
2
ON/OFF [Fig. 2] using
the up/down keys (
△▽
), and press the enter key (ENTER).
▶
The left side of the display changes to the O
2
ON/OFF screen, and
the current set value (ON) is displayed [Fig. 3].
4. Use the up/down keys (
△▽
) to change the control setting value.
Each time the up/down key is pressed the value changes between ON
and OFF.
5. Press the enter key (ENTER).
▶
If it is set to ON: the value is confirmed and the display changes to the
O
2
Setting screen.
If it is set to OFF: the value is confirmed and the display changes to
the Set Value screen.
6. Press the menu key (MENU/HOME) to display the home screen.
The display returns to the home screen automatically when 90
seconds has passed without any key operation.
Set Value
Temperature
:
37.0
o
C
CO
2
ON/OFF
: ON
CO
2
Setting
: 5.0%
O
2
ON/OFF
: ON
O
2
Setting
: 5.0%
Overtemp
: 53.0
o
C