68
ALARM PARAMETERS
Setting ring back delay
This function reactivates the alarm when the alarm state still continues and preset amount of time has
elapsed after the BUZZER STOP key was pressed to stop the alarm buzzer.
Settable range: 0 – 99 minutes
(when the setting is 0 minutes, the Ring Back Delay becomes
“OFF.”)
Factory setting: 30 minutes
1. On the home screen, press the menu key (MENU/HOME).
▶
The left side of the display changes to the Menu screen.
2. On the Menu screen, move the cursor to Alarms & Controls [Fig. 1]
using the up/down keys (
△▽
), and press the enter key (ENTER).
▶
The display changes to the Alarms & Controls screen.
3. On the Alarms & Controls screen, move the cursor to Other Alarm Set
[Fig. 2] using the up/down keys (
△▽
), and press the enter key (ENTER).
▶
The display changes to the Other Alarm Set screen.
4. On the Other Alarm Set screen, move the cursor to Ring Back Delay
[Fig. 3] using the up/down keys (
△▽
), and press the enter key (ENTER).
▶
The right side of the display changes to the Ring Back Delay screen,
and the current setting value (30) is displayed [Fig. 4].
5. Move the cursor using the left/right keys (
◁▷
) and select the digit you
want to change, and then use the up/down keys (
△▽
) to change the
number.
6. Press the enter key (ENTER).
▶
The value is confirmed and the display returns to the Other Alarm Set
screen.
7. Press the menu key (MENU/HOME) to display the home screen.
The display returns to the home screen automatically when 90
seconds has passed without any key operation.
Note:
In the case of Err01 (CO
2
gas cylinder empty), Err11 and Err12
(CO
2
sensor error) or Door alarm, the alarm does not reactivate because
the alarm itself is canceled by pressing the BUZZER STOP key. Also, in
the case of Err18 (UV lamp failure), the alarm does not reactivate if the
BUZZER STOP key is pressed after the UV lamp ON period is over
(refer to page 90).
<Left side of the display> [Fig. 1]
Menu
Set Value
Data Log
Lock
Alarms & Controls
<Right side of the display> [Fig. 4]
Ring Back Delay
30
min
0min – 99min
<Left side of the display> [Fig. 2]
Alarms & Controls
Temp/Gas Alarm Set
Other Alarm Set
UV Setting
Gas Supply Setting
Date & Time
Keypad Click Set
▼
<Left side of the display> [Fig. 3]
Other Alarm Set
Door Alarm Delay
Ring Back Delay
Remote Alarm