73
UV LAMP (OPTION)
Automatic extension of UV lamp ON period
The amount of UV irradiation decreases as the total UV lamp ON period increases. To compensate for this,
UV lamp ON period is automatically extended as the total UV lamp ON period increases (the value set for
UV Lighting Time is not changed).
Settable range: 0% – 40% (cannot be set)
Example) UV Lighting Time: 10 minutes plus Auto-Extended Time: 10 %
→
UV lamp lights for 11 minutes.
1. On the home screen, press the menu key (MENU/HOME).
▶
The left side of the display changes to the Menu screen.
2. On the Menu screen, move the cursor to Alarms & Controls [Fig. 1]
using the up/down keys (
△▽
), and press the enter key (ENTER).
▶
The display changes to the Alarms & Controls screen.
3. On the Alarms & Controls screen, move the cursor to UV Setting
[Fig. 2] using the up/down keys (
△▽
), and press the enter key
(ENTER).
▶
The display changes to the UV Setting screen.
4. On the UV Setting screen, move the cursor to Auto-Extended Time
(Fig. 3) using the up/down keys (
△▽
), and press the enter key
(ENTER).
▶
The right side of the display changes to the Auto-Extended Time
screen, and the current UV lighting time's automatic extension ratio is
displayed [Fig. 4].
5. Press the enter key (ENTER).
▶
The display returns to the UV Setting screen.
6. Press the menu key (MENU/HOME) to display the home screen.
The display returns to the home screen automatically when 90
seconds has passed without any key operation.
<Left side of the display> [Fig. 2]
Alarms & Controls
Temp/Gas Alarm Set
Other Alarm Set
UV Setting
Gas Supply Setting
Date & Time
Keypad Click Set
▼
<Left side of the display> [Fig. 3]
UV Setting
UV Lighting Time
UV Life Counter
Auto-Extended Time
UV 24h Mode Start
Auto-Extended Time
10
%
<Right side of the display> [Fig. 4]
<Left side of the display> [Fig. 1]
Menu
Set Value
Data Log
Lock
Alarms & Controls