201
5-0
3
46
Pulscon LTC57 PROFIBUS PA
Electrical connection
7.2
Connecting the device
Required tools and accessories:
• For instruments with safety pin for the lid: AF3 Allen key
• Wire stripping pliers
• When using stranded wires: wire end sleeves.
Figure 7.2
Connecting the device
1. Loosen the screw of the securing clamp of the connection compartment cover and turn the
clamp 90
°
counterclockwise.
2. Unscrew the connection compartment cover.
3. Push the cable through the cable entry. To ensure tight sealing, do not remove the sealing
ring from the cable entry.
4. Strip the cable.
5. Strip the cable ends 10 mm (0.4 in). For stranded cables, also attach wire end ferrules.
6. Firmly tighten the cable glands.
Warning!
Explosion hazard!
• Comply with the relevant national standards.
• Observe the specifications in the Safety Instructions (SI).
• Only use the specified cable glands.
• Check whether the supply voltage matches the specifications on the nameplate.
• Before connecting the device: Switch the supply voltage off.
• Before switching on the supply voltage: Connect the potential bonding line to the exterior
ground terminal.
1.
2.
3.
10 (0.4)
10 (0.4)
mm (in)
4.
5.
5.
Содержание Pulscon LTC57