Pulscon LTC57 PROFIBUS PA
Mounting
6.1.2
Applications with restricted mounting space
Mounting with remote sensor
The device version with a remote sensor is suited for applications with restricted mounting
space. In this case the electronics housing is mounted at a separate position from which it is
easier accessible.
Figure 6.2
• Product structure, feature "Probe design": option B "Sensor remote, 3 m/9 ft cable"
• The remote cable is supplied with these device versions, minimum bending radius:
100 mm (4 in)
• A mounting bracket for the electronics housing is supplied with these device versions.
Mounting options:
• Wall mounting
•
Pipe mounting; diameter: 42 to 60 mm (1-1/4 to 2 in)
• The connection cable has got one straight and one angled plug (90
°
). Depending on the
local conditions the angled plug can be connected at the probe or at the electronics
housing.
A
Angled plug at the probe
B
Angled plug at the electronics housing
C
Length of the remote cable as ordered
A
B
6 Nm
(4.42 lbf ft)
6 Nm
(4.42 lbf ft)
6 Nm
(4.42 lbf ft)
C
C
6 Nm
(4.42 lbf ft)
r
= 100 (4)
min
r
= 100 (4)
min
Note!
Probe, electronics and connection cable are adjusted to match each other. They are marked
by a common serial number. Only components with the same serial number shall be
connected to each other.