GEA RM1
67
Appendix
TEST circuit diagram
Wiring diagram for GEORESISTIVITYMETER BOX TESTING
FIGURE 49 - ELECTRICAL DIAGRAM FOR TESTING BOX
Performing the measurement (after connecting A, B, M, N to the respective sockets on
the Resistivity Meter and the energizer to its socket) the ratio ΔV / I must be equal to the
value of R2 then 15 Ω (± 10%)
CAUTION:
the tensions produced by the energizer are potentially lethal!
Содержание GEA RM1 1.0
Страница 4: ...4 TEST circuit diagram 67...
Страница 5: ......
Страница 13: ...GEA RM1 9 How to use FIGURE 4 EXAMPLE OF INTEGRAL TABLE FIGURE 5 EXAMPLE WITH SIMPLIFIED VIEW...
Страница 15: ...GEA RM1 11 How to use FIGURE 8 FIGURE 7...
Страница 16: ...12 FIGURE 9 MEASURE WITH INCREASED MN FIGURE 10 REPEAT MEASURE...
Страница 17: ...GEA RM1 13 How to use FIGURE 11 EXECUTE MODIFY REPEAT FIGURE 12 MEASURING EXAMPLE WITH SIGMA 5...
Страница 50: ...46 FIGURE 36 INITIAL PAGE WITH DISCONNECTED SYSTEM FIGURE 37 INITIAL PAGE WITH CONNECTED SYSTEM AND SWITCHED ON...
Страница 56: ...52 FIGURE 42 CONSTRUCTION OF A MANUAL TABLE FIRST PART FIGURE 41 CONSTRUCTION OF A MANUAL TABLE SECOND PART...