Description of the systems interface in standalone mode
-
PS: Spontaneous Potential, automatically measured by the system and used to
purify the final measurement of the resistance, in Volts.
-
IP: decay time measured in milliseconds.
-
K: geometric constant dependent on the relative positions of the electrodes used
for the measurement and used for the calculation of Rho. It is calculated by the
application in the cases in which measurement campaigns supported by tables are
carried out, it is not possible to calculate it in the case of a single measurement in
which the positions of the electrodes are not known. In the second case the value
of K is not shown on the display.
-
R = V / I: Resistance measured as a ratio between V (M-N) and I (A-B), in Ohms.
Measurement of spontaneous potential
Spontaneous Potential measurement is performed as a single measure, the result of which
is presented to the operator on the video as shown in Figure 19.
Below are listed the values done after a measure of only spontaneous potential.
- PS: Spontaneous Potential, automatically measured by the system, in Volts.
If the measurement is made as an effective single measurement and without the aid of a
measurement session, the operator must copy the electrical parameters obtained with the
measurement, if deemed necessary, as the single measurement mode does not allow the
measurement memory storage of the result tool.
In case the measurement execution is controlled through the execution of a measurement
Execution, editing and repetition of a measure in session
) after
visualizing the measurement result, the operator must accept or reject the measurement
result on the instrument. In the first case the measured electrical parameters will be
F
IGURE
31
-
RESULTS OF A SPONTANEOUS POTENTIAL
SINGLE MEASURE
Содержание GEA RM1 1.0
Страница 4: ...4 TEST circuit diagram 67...
Страница 5: ......
Страница 13: ...GEA RM1 9 How to use FIGURE 4 EXAMPLE OF INTEGRAL TABLE FIGURE 5 EXAMPLE WITH SIMPLIFIED VIEW...
Страница 15: ...GEA RM1 11 How to use FIGURE 8 FIGURE 7...
Страница 16: ...12 FIGURE 9 MEASURE WITH INCREASED MN FIGURE 10 REPEAT MEASURE...
Страница 17: ...GEA RM1 13 How to use FIGURE 11 EXECUTE MODIFY REPEAT FIGURE 12 MEASURING EXAMPLE WITH SIGMA 5...
Страница 50: ...46 FIGURE 36 INITIAL PAGE WITH DISCONNECTED SYSTEM FIGURE 37 INITIAL PAGE WITH CONNECTED SYSTEM AND SWITCHED ON...
Страница 56: ...52 FIGURE 42 CONSTRUCTION OF A MANUAL TABLE FIRST PART FIGURE 41 CONSTRUCTION OF A MANUAL TABLE SECOND PART...