ANUP5204 Aicure
Irradiation Modes
21
UV Autocontrol Feature
This feature compensates for loss of UV intensity due to lamp wear. This maintains stable UV intensity
until the end of the lamp lifetime.
To begin automatic lamp output compensation, on the "AUTO CONTROL" section of the operation
panel, press the "ON/OFF" button until you hear a beep (at least 3 seconds). During compensation,
the LED above the "ON/OFF" button is lit.
To end UV auto-control, press the "ON/OFF" button again until you hear a beep (at least 3 seconds).
Automatic lamp output compensation ends, and the LED goes off.
Setting Lock Feature
This feature allows you to lock your settings, so they will not be accidentally changed. On the
operation panel, simultaneously press and hold the up and down arrow buttons for at least three
seconds. The value becomes locked and cannot be changed.
To release the lock on the setting, press the up and down arrow buttons simultaneously again for
three seconds.
Counter Feature
A count is kept of the number of times the shutter is opened (up to 9,999). On the operation panel,
press the "COUNTER" button. The count appears on the display (the LED above the button lights).
To reset the counter, while the counter value is being displayed, simultaneously press and hold the
"COUNTER" and "SET" buttons until you hear a beep (at least 1 second).
If you need to modify the counter value, for instance to cancel out test operation cycles, then while the
counter value is being displayed, press and hold the up and down arrow buttons. The value is
changed by one at a time. Once the value you desire has been reached, press and hold the "SET"
button for at least one second. Note, however, that the value can only be changed while the shutter is
closed.
1. The UV Autocontrol feature is retained even if the power is interrupted.
2. Even when you are not using the UV Autocontrol feature, it is keeping a record of
elapsed time; even if you cut into autocontrol midway, it performs compensation.
3. The UV Autocontrol feature is based on elapsed time. When replacing a lamp, you
must install the new lamp, then reset the cumulative lamp use time (see p. 38).