HygroPro
II
User's Manual
63
Appendix A. HygroPro
II
HART
®
Field Device Specification
A.9.24 Command 163 (0xA3): Read Temperature Coefficients
This command reads the temperature coefficients.
Request Data Bytes
Response Data Bytes
Command-Specific Response Codes
A.9.25 Command 166 (0xA6): Set Saturation Table Index to Read Table Array Element
Set the Saturation table index for the reading of an array element.
Request Data Bytes
Response Data Bytes
5
Error
Too Few Data Bytes Received
6
Error
Device-Specific Command Error
7
Error
In Write Protect Mode
8-15
Undefined
16
Error
Access Restricted
8-127
Undefined
Byte
Format
Description
None
Byte
Format
Description
0 ~ 3
Float
Calibration Temperature
4 ~ 7
Float
Temperature Dependence
8 ~ 11
Float
Dampen Factor
Code
Class
Description
0
Success
No Command-Specific Errors
1-127
Undefined
Byte
Format
Description
0
Unsigned-8
Index of array element to be read
Byte
Format
Description
0
Unsigned-8
Index of array element to be read
Code
Class
Description
Содержание HydroProII
Страница 1: ...panametrics com 910 331 A November 2022 HygroProII Moisture Transmitter User s Manual...
Страница 2: ......
Страница 4: ...ii no content intended for this page...
Страница 12: ...Preface x HygroProII User s Manual no content intended for this page...
Страница 42: ...Chapter 2 Operation 30 HygroProII User s Manual no content intended for this page...
Страница 52: ...Chapter 4 Specifications 40 HygroProII User s Manual no content intended for this page...
Страница 100: ...Appendix B HygroProII Field Service Update User Manual 88 HygroProII User s Manual no content intended for this page...
Страница 102: ...Warranty 90 HygroProII User s Manual no content intended for this page...
Страница 103: ......