Chapter 2. Operation
The main menu map shown in two sections above can be accessed by pressing the Escape
key. In Figure 16,
red
text
indicates special inputs required to access the succeeding section of the menu map, while
green text
indicates
the “default” values of the particular menu map branch.
2.1.1
Keypad
After entering the setup program, the keys on the HygroPro
II
keypad (see Figure 15 on page 17) perform the following
functions:
• Enter
– confirm a selection or move to the next screen
• Escape
– cancel a selection, move to the previous screen or open main menu
• Up
⬆
– scroll upward through a list of options or increase the valux‘e of a selected character
• Down
⬇
– scroll downward through a list of options or decrease the value of a selected character
• Left
⬅
– move the cursor to the next character/field to the left
• Right
➡
– move the cursor to the next character/field to the right
2.2 Basic Setup
The HygroPro
II
transmitter is easily programmed to meet the user’s requirements by referring to the menu map in
Figure 16 on the previous page and following the instructions in this section. As noted in the figure, the following six
top-level parameters can be accessed and set on the HMI without any passcode-
a) No. of Views
b) Display
c) Output
d) HART
e) Constants
f) About
The
Settings menu requires a 6-digit User passcode (default value 111111) while the Service menu (see Chapter 3)
requires a device-specific 6-digit Service passcode (contact Panametrics Tech Support) for access.
2.2.1
Calibration Reminder settings on initial power up
One of the new features offered by the HygroPro
II
transmitter is automated calibration reminders to the end user.
This feature uses an internal clock to determine elapsed time from the HygroRTE probe calibration date.
The initial power up screen on the HygroPro
II
is shown on the left. After
going through the boot up splash screen, the device displays the
following three choices for the calibration interval:
• 1 Year
• 6 Months
• Skip
The Calibration Interval is defined in months from the probe calibration
date after it has been read from the probe non-volatile memory upon
initial power up. The end user is expected to set this calibration interval
depending of their facility’s policy to one of the above settings.
CAUTION!
Panametrics strongly recommends that the HygroRTE
probe be sent back to the Billerica, USA or Shannon, IE
facilities for recalibration every 6–12 months to ensure
accuracy and reliability of the moisture readings.
Содержание HydroProII
Страница 1: ...panametrics com 910 331 A November 2022 HygroProII Moisture Transmitter User s Manual...
Страница 2: ......
Страница 4: ...ii no content intended for this page...
Страница 12: ...Preface x HygroProII User s Manual no content intended for this page...
Страница 42: ...Chapter 2 Operation 30 HygroProII User s Manual no content intended for this page...
Страница 52: ...Chapter 4 Specifications 40 HygroProII User s Manual no content intended for this page...
Страница 100: ...Appendix B HygroProII Field Service Update User Manual 88 HygroProII User s Manual no content intended for this page...
Страница 102: ...Warranty 90 HygroProII User s Manual no content intended for this page...
Страница 103: ......