HygroPro
II
User's Manual
65
Appendix A. HygroPro
II
HART
®
Field Device Specification
A.9.28 Command 169 (0xA9): Read Hygro Calibration Curve Table Index to Read Table Array
Element
Read the Hygro Calibration index for the reading of an array element.
Request Data Bytes
Response Data Bytes
Command-Specific Response Codes
A.9.29 Command 170 (0xAA): Read Loop Measurement Type
This command is to read the loop measurement type to 1 of 3 enumerations:
0 – 4-20 mA, 1 – NAMUR, 2 – Special Zero and Span Values
When the loop measurement type is set to Special, then the Special span value and special zero value will determine
the endpoint mA values instead of 4 mA and 20 mA.
Request Data Bytes
Response Data Bytes
Command-Specific Response Codes
Byte
Format
Description
0
Byte
Format
Description
0
Unsigned-8
Index of array element to be read
Code
Class
Description
0
Success
No Command-Specific Errors
1-127
Undefined
Byte
Format
Description
0
Byte
Format
Description
0
Unsigned-8
Loop Measurement Type
1~4
Float
Special Zero Value
5~8
Float
Special Span Value
Code
Class
Description
0
Success
No Command-Specific Errors
1-4
Undefined
5
Error
Too Few Data Bytes Received
Содержание HydroProII
Страница 1: ...panametrics com 910 331 A November 2022 HygroProII Moisture Transmitter User s Manual...
Страница 2: ......
Страница 4: ...ii no content intended for this page...
Страница 12: ...Preface x HygroProII User s Manual no content intended for this page...
Страница 42: ...Chapter 2 Operation 30 HygroProII User s Manual no content intended for this page...
Страница 52: ...Chapter 4 Specifications 40 HygroProII User s Manual no content intended for this page...
Страница 100: ...Appendix B HygroProII Field Service Update User Manual 88 HygroProII User s Manual no content intended for this page...
Страница 102: ...Warranty 90 HygroProII User s Manual no content intended for this page...
Страница 103: ......